ref: 01ESHI98121 sku: HI98121
  • HANNA intruments pocket pH, ORP and temperature meter
  • Backlit display for easy readability
  • Compact, robust
  • Fully watertight, floating housing
  • Easily replaceable electrodes
  • 2-point ph calibration and ORP calibration is factory calibrated
€244.00
Free shipping
Free Shipping + info

Technical characteristics

Tester for measuring pH, ORP and temperature.

Meter with a two-level display that allows you to view the temperature at all times.

The pH electrode consists of a renewable fibre junction that can be removed quickly and easily for replacement using a key supplied with the instrument.

The temperature sensor, housed in a metal tube, maintains fast response times thanks to its external position.

The instrument is powered by common batteries. When the instrument is switched on, it indicates the percentage of battery life remaining, so that its operating status can be monitored.

The measurement stability indicator allows reliable readings to be obtained.

Designed with ergonomic styling, it is capable of floating, making it easy to retrieve in the event of an accidental fall into the water.

Specifications Hanna HI98121:

ORP range. +/- 1000mV.
PH range. -2.00 to 16.00 pH range.
Temperature range. from -5 to 60ºC / from 23 to 140ºF.
ORP resolution. 1 mV.
PH resolution. 0.01 pH.
Temperature resolution. 0.1ºC/0.1ºF.
Accuracy (at 20ºC) ORP. +/- 2 mV.
Accuracy (at 20ºC) pH +/- 0.05 pH.
Accuracy (at 20ºC) temperature. +/-0.5ºC +/- 1ºF.
ORP calibration. factory pre-calibrated
PH calibration. automatic in 1 or 2 points with 2 sets of memorised buffers (pH 4.01/7.01/10.01 or 4.01/6.86/9.18).
Temperature compensation. automatic per pH.
Electrodes. hI 73127 replaceable pH electrode (included) (fixed ORP electrode).
Battery type / duration.

4 x 1.5V with BEPS / approx. 250 hours continuous use.
auto-off after 8 minutes of inactivity.

Working conditions. from -5 to 50°C; max. 100 % R.H.
Dimensions. 163 x 40 x 26 mm.
Weight. 100 g.

Supplied with:

  • Protective cover.
  • Spanner for electrode replacement.
  • Batteries.
  • Instructions.